How Cell Culture Medium Can Decrease Cell Viability During A Flow Cytometry Cell Sorting Experiment

When setting up a cell sorting experiment, there are many things to consider.

You must consider which controls you’re going to use, which quality control beads are ideal, how you’re going to compensate the experiment, which instrument and which instrument settings are ideal, on and on to how you plan to analyze, gate, and present your data.

With so many things to consider, it’s easy to lose site of the small things that can drastically affect the viability of your cells, including the composition of your suspension buffer. 

The composition of the suspension buffer for preparation, staining, analyzing and sorting is perhaps the most important parameter for maintaining viability during a cell sorting experiment.

While the precise components of a buffer can differ depending on the cell type, there are 5 key points to keep in mind…

1. Cell culture medium is usually not a good choice. 

Most cell culture media are formulated with a CO2-carbonate buffering system that is optimized for tissue culture carbon dioxide partial pressure, which is higher than the typical ambient carbon dioxide partial pressure.

When these kinds of buffers are left in typical atmospheric conditions, carbon dioxide will evaporate from the medium and cause the pH of the medium to rise into the alkaline range (in phenol red-containing formulations, the color will become more purple), which expectedly can significantly affect viability.

While tissue culture medium is often used as a suspension buffer without deleterious effect, longer cell sorts and the sensitivity of the cell type can certainly exacerbate the effect of pH on viability.

Additionally, although it’s important to formulate the buffer with some kind of protein, tissue culture media may contain serum at concentrations of upwards of 10-15%, which can be deposited over time on the sample line and cause flow rate irregularities during the sort.

If tissue culture medium must be used, a HEPES-based buffering system, which is not dependent on carbon dioxide, is a preferable and safer choice.

2. Consider adding protein to your buffer.

PBS or HBSS can work well as a base for the suspension fluid, but you should add some kind of protein, typically 1-2% BSA or FBS, to keep the cells happy.

Incidentally, HBSS is formulated with a CO2-carboante buffering system, but this system is meant to work at atmospheric/ambient CO2 partial pressure rather than tissue-culture CO2 partial pressure.

3. Add EDTA to prevent clumping.

EDTA in the suspension buffer, at 1-5 mM, can be a panacea for samples that tend to form aggregates and clog the instrument. This molecule interferes with cell-cell adhesion molecules by sequestering the divalent cations that these molecules require to function.

Therefore, it can greatly reduce the incidence of clumping in the sample. Preventing clogs has an indirect effect on viability—clogs cause downtime during the sort, requiring cells to remain in suboptimal conditions for longer than necessary. The faster the cells are manipulated, the lower the likelihood that they will succumb to apoptosis or necrosis.

If you will be using DNAse (see #4 below), then EGTA is a better choice for chelation to prevent clumping. EDTA will chelate Mg++ ions, which are necessary for DNAse, while EGTA has a much lower affinity for Mg++.

4. Add DNAse to deal with dead cells.

Certain kinds of samples, especially those harvested from some kinds of solid tissues, will inevitably contain a high proportion of dead cells. If this is the case, it may be advantageous to add DNAse to the suspension buffer used for the sorting process.

Dead cells will release genomic DNA into the buffer, which tends to aggregate and form clumps. These clumps can trap cells, cause clogs and result in an overall drop in recovery after the sort. Recommended concentrations vary, but 200 µg/mL has been reported in Current Protocols In Cytometry to be effective. Alternatively, debris from dead cells can be removed during the preparation either through centrifugation gradients or depletion via magnetic separation.

5. Keep everything at the right temperature.

The temperature at which the sample is maintained during preparation, staining, and sorting is another important parameter to keep cells alive. The proper temperature strongly depends on the cell type and application.

While many researchers have success preparing and sorting cells at temperatures around 0°C, this may not be the ideal temperature for every cell type. 

Nevertheless, there are some general benefits for keeping the cells on ice. Most importantly, a low temperature will slow down metabolic activities and will prevent biochemical or molecular changes that may affect cytometric or downstream results.

For example, some cells, under certain conditions, may shed or internalize antibody-bound cell surface receptors, resulting in diminished staining and resolution on the cytometer. Keeping cells cold can significantly reduce this effect.

Additionally, under the stress of preparation, staining, and the sorting process, some cells may undergo molecular changes—gene expression or otherwise—that could affect downstream experiments or analyses. Keeping cells cold will slow down activity and possibly mitigate any transcriptional effects brought about by the experiment.

Finally, stressed cells may begin to apoptose, lowering your recovery. By keeping your cells cold, you may slow down or mitigate this process so they can be deposited in the collection buffer before progressing further through this pathway.

When designing your flow cytometry experiment, check the literature to see if any particular suspension buffer and temperature conditions have been successful historically in other protocols for the particular cell type you are working. Buffer with protein, EDTA, and DNAse is often a better choice than using your cell culture medium alone. In terms of temperature—cold, usually 0°C, is a good place to start, but temperature can be surprisingly cell-type dependent, so make sure you do your research.

To learn more about current best practices in flow cytometry and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class
Tim Bushnell, PhD
Tim Bushnell, PhD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Similar Articles

Which Fluorophores To Use For Your Microscopy Experiment

Which Fluorophores To Use For Your Microscopy Experiment

By: Heather Brown-Harding, PhD

Fluorophore selection is important. I have often been asked by my facility users which fluorophore is best suited for their experiments. The answer to this is mostly dependent on whether they are using a widefield microscope with set excitation/emission cubes or a laser based system that lets you select the laser and the emission window. Once you have narrowed down which fluorophores you can excite and collect the correct emission, you can further refine the specific fluorophore that is best for your experiment.  In this blog  we will discuss how to determine what can work with your microscope, and how…

4 No Cost Ways To Improve Your Microscopy Image Quality

4 No Cost Ways To Improve Your Microscopy Image Quality

By: Heather Brown-Harding, PhD

Image quality is critical for accurate and reproducible data. Many people get stuck on the magnification of the objective or on using a confocal instead of a widefield microscope. There are several other factors that affect the image quality such as the numerical aperture of the objective, the signal-to-noise ratio of the system, or the brightness of the sample.  Numerical aperture is the ability of an objective to collect light from a sample, but it contributes to two key formulas that will affect your image quality. The first is the theoretical resolution of the objective. It is expressed with the…

What Is Total Internal Reflection Fluorescence (TIRF) Microscopy & Is It Right For You?

What Is Total Internal Reflection Fluorescence (TIRF) Microscopy & Is It Right For You?

By: Heather Brown-Harding, PhD

TIRF is not as common as other microscopy based techniques due to certain restrictions. We will discuss these restrictions, then analyze why it might be perfect for your experiment.  TIRF relies on an evanescent wave, created through a critical angle of coherent light (i.e. laser) that reaches a refractive index mismatch.  What does it mean in practice?  A high angle laser reflects off the interface of the coverslip and the sample. Although the depth that this wave penetrates is dependent on the wavelength of the light, in practice it is approximately 50-300nm from the coverslip. Therefore, the cell membrane is…

5 Drool Worthy Imaging Advances Of 2020

5 Drool Worthy Imaging Advances Of 2020

By: Heather Brown-Harding, PhD

2020 was a difficult year for many, with their own research being interrupted- either by lab shutdowns or recruitment into the race against COVID-19. Despite the challenges, scientists have continued to be creative and have pushed the boundaries of what is possible. These are the techniques and technologies that every microscopist was envious of in 2020. Spatially Resolved Transcriptomics Nature Methods declared that spatially resolved transcriptomics was the 2020 method of the year. These are a  group of methods that combine gene expression with their physical location. Single-cell RNA sequencing (scRNAseq) was originally developed for cells that had been dissociated…

Picking The Right Functional Imaging Probe

Picking The Right Functional Imaging Probe

By: Heather Brown-Harding, PhD

As biologists, we study the process of life, however, it’s intricacies cannot be captured by a snapshot in time. Generally, the easiest imaging experiments are those where the samples are stained, fixed, and imaged within a few days of procurement, but that too doesn’t capture the dynamic processes common in cells and organisms. Live cell imaging when combined with reporters serves as a powerful tool to provide solid imaging data. Cameleon —one of the first reporters— was developed in 1997 in Roger Tsien’s lab.  Cameleon is a green fluorescent protein (GFP) that undergoes a conformational change in the presence of…

6 Areas Of Consideration For Flow Cytometry Cell Cycle Analysis

6 Areas Of Consideration For Flow Cytometry Cell Cycle Analysis

By: Tim Bushnell, PhD

Cell cycle seems like such a straightforward assay. At its heart, it is a one-color assay and should be a simple protocol to follow. However, as discussed before, fixation and dye concentrations are critical. Once those are optimized, it becomes important to run the cells low and slow in order to get the best quality histograms for analysis — the topic of another blog. Adding the critical CEN and TEN controls will help standardize the assay, and ensure consistency and reproducibility between runs while helping identify non-standard (aneuploid, polyploid) populations from normal ploidy. Trying to isolate and focus on specific…

Why Cell Cycle Analysis Details Are Critical In Flow Cytometry

Why Cell Cycle Analysis Details Are Critical In Flow Cytometry

By: Tim Bushnell, PhD

Cell cycle analysis appears to be deceptively easy in concept, but details are absolutely critical. It is not possible to hide the data if there is poor sample preparation, incorrect dye ratios, too much (or too little) staining time, etc. Forgetting RNAse when using PI will doom your data to failure. Take these basics into account as you move into performing this simple, yet amazingly informative assay.

How To Choose The Correct Antibody For Accurate Flow Cytometry Results

How To Choose The Correct Antibody For Accurate Flow Cytometry Results

By: Tim Bushnell, PhD

With the added emphasis on reproducibility, it is critical to look at every step where experiments can be improved. No single step makes an experiment more reproducible, rather it is a process, making changes at each stage that leads to reproducibility. Antibodies comprise a critical component that needs to be reviewed. As Bradbury et al. in a commentary in Nature pointed out, the global spending on antibodies is about $1.6 billion a year, and it is estimated about half of that money is spent on “bad” antibodies. This does not include the additional costs of wasted time and effort by…

5 Essential Beads For Flow Cytometry Experiments

5 Essential Beads For Flow Cytometry Experiments

By: Tim Bushnell, PhD

Flow cytometry is designed to measure physical and biochemical characteristics of cells and cell-like particles using fluorescence. Fundamentally, any single-particle suspension (within a defined size range) can pass through the flow cytometer. Beads, for better or worse, are a sine qua non for the flow cytometrist. From quality control,to standardization, to compensation, there is a bead for every job. They are important — critical, even — for flow cytometry.

Top Technical Training eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.