What Is Flow Cytometry Light Scatter And How Cell Size And Particle Size Affects It

Light scatter is unfortunately one of the more misunderstood concepts in cytometry.

When you first learn flow cytometry, you were likely told the misleading phrase, “Forward scatter signal intensity is proportional to cell size, and side scatter signal intensity is proportional to cell granularity.”

The short story here is that, yes, this can be the case when we’re dealing with typical cells (consider blood granulocytes, which are bigger than lymphocytes, have more intense FSC signals; granulocytes are more “granular” than lymphocytes and thus have more intense SSC signals). However, the truth is that light scatter complexity is belied by this perennial but unsatisfactory introductory explanation.

While particle size (particle radius) certainly does influence light scatter signal, its intensity is a function of a combination of factors, including:

  •   wavelength of laser illumination
  •   collection angle, and
  •   refractive index of the particle and flow medium (sheath)

It turns out that light scatter intensity has a strong dependence on the relationship between the size of the particle to the wavelength of the laser. More specifically, particles with diameters that are larger than the wavelength of the laser will scatter light with a different pattern than particles that are smaller than the wavelength of the laser.

Keep in mind that we most typically measure scatter using 488 nm excitation (and sometimes 405 nm excitation), so particles or cells with diameters larger than about 0.5 microns will behave differently than particles with diameters significantly smaller than 0.5 microns.

It turns out that there is a physical theory, named after the German physicist Gustav Mie (1869-1957), that predicts and explains the behavior of light scattering of particles larger than the wavelength of illumination. Essentially, Mie’s theory predicts that the intensity of scattered light has a strong angular dependence.

In other words, the intensity of the signal generated from scattered light depends on the angle at which we collect and direct the light towards a detector. Mie’s theory explains why we use a forward scatter (FSC) detector to measure the light scatter signal of typical mammalian cells, which are much bigger than the wavelength of the illumination source (typically the 488 nm laser).

How Small Particles Affect Forward Scatter

Forward scatter detectors collect light at small angles relative to the incident beam and can take advantage of the fact that cells preferentially scatter light in this “forward” direction.

Additionally, because cells scatter so much light in the forward direction, we can save money and use a less sensitive detector to measure this light. As such, forward scattered light is traditionally and often effectively measured with a photodiode, rather than the more sensitive photomultiplier used to measure fluorescence and side scatter.

However, the story is quite different for particles that are smaller than the illumination wavelength (<488 nm or <~0.5 um), like microvesicles and ectosomesIt turns out that light scatter by particles of this size range is NOT dependent on the angle at which it is measured. This has very significant implications for small particle analysis. Most importantly, because small particles do not preferentially scatter light in the forward direction as do cells, the resolution in this detector may not be sufficient to measure or even identify these particles above background.

How Small Particles Affect Side Scatter

Side scatter, given its orientation in the quieter fluorescence collection path, as well as its traditional detection by the much more sensitive photomultiplier tube, will likely have much better resolution than forward scatter.

Similarly, we often use side scatter as the “trigger” or “threshold” parameter when measuring small things like bacteria, microparticles, or microvesicles for the same reason—better sensitivity allows us to better distinguish and measure small particles above background.

Interestingly, scatter gets dim VERY quickly when particles have diameters below the wavelength of illuminating light, considering that scatter intensity decreases with a dependence on r6 of the particle. The bottom line is that small things like extracellular vesicles can be incredibly difficult to detect using scatter signals. This can make publishing flow cytometry data on small particles very difficult.

What Is An Obscuration Bar?

Instrument manufacturers and operators often take advantage of this property of small particle light scatter by installing an adjustable obscuration bar on the forward scatter detector.

The forward scatter obscuration bar is a universal component of this detector that helps to diminish background in the FSC detector by blocking laser light from interacting with the detector. When no particle is present in the laser beam, scattered laser light hits and is blocked by the obscuration bar. On the other hand, when a particle is present in the laser beam, laser light refracted (scattered) by the particle passes over the bar and triggers signal associated with that particle.

An adjustable obscuration bar can be rotated to expose a wider or narrower surface to the laser beam, blocking more or less laser light, respectively, from hitting the detector. To resolve smaller particles from optical noise, it can be helpful to block more laser light from hitting the detector, which can be accomplished by widening the bar.

Moreover, because small particles do not preferentially scatter light in the forward direction, the proportion of signal blocked by the bar of the total signal is less significant than it would be for larger particles, which do preferentially scatter in the forward direction. This strategy is most effective when a photomultiplier tube is used for forward scatter detection than a photodiode, given that the former is much more sensitive than the latter.

This relationship between particle size, wavelength of illumination, and scatter angle can also help explain the side scatter properties of cells. Typical cellular side scatter signal is much more robustly correlated to granularity than forward scatter is to cell size. In fact, the cytoplasmic “granules” that influence side scatter signal are often smaller than 0.5 um and will thus scatter in a non-Mie pattern. Side scatter of mammalian cells and side scatter of small particles are not terribly different after all.

What Is A Refraction Index?

In addition to the wavelength of the laser and the collection angle of the scatter optics, another factor that significantly contributes to the light scattering intensity of a particle is the refractive index of the suspension medium (sheath, which is essentially water) and the particle itself.

For particles that obey Mie’s predictions, light scattering is largely composed of laser light refraction. When a particle or cell is absent at the interrogation point, the laser’s intersection with the particle or cell produces a characteristic refraction, the “ring of diffraction,” that we are most likely acquainted with on sense-in-air cell sorters. This ring of light spreads outwards from the stream in all directions, in the same plane as the laser beam, and is blocked from entering the forward and side scatter detection paths by the obscuration bars in front of each.

However, the presence of a cell in the laser beam changes the composition of the medium through which the light travels. Laser light now passes through the cytoplasm—which contains protein, lipids, and carbohydrates rather than simply the water it passed through in the absence of a cell—which therefore causes the light to bend differently than it does when it passes through the sheath fluid alone. This refraction of light by a cell is a function of the difference in refractive indices (RI) between the media through which light passes, and it causes laser light to bend in such a way that it passes over the obscuration bar and interacts with the detector, generating scatter signal.

Every type of material has an associated refractive index. When light passes from one material to another—say, from the saline of the sheath fluid to the material of the cell and then back through the sheath fluid again—the amount of light that bends due to this transition is proportional to the difference between the refractive indices between the media. The bigger the difference, the more the scatter.

Why Beads Are Not Good For Calibrating Particles

It is precisely this property of light scatter that makes it a terribly unreliable measurement of cell size.

Two particles or cells of exactly the same size may have different refractive indices, due to their composition (e.g. cytoplasmic proteins), and will therefore generate scatter signals with different intensities. This is also precisely why synthetic beads are also terrible size calibrators.

The refractive index of a typical polystyrene bead (1.59) can be significantly different than a cell’s, resulting in very different scatter signals between a bead and a cell of the same diameter. Given the fact that cells are composed of much more water than a polystyrene bead is and that water’s RI is 1.333, polystyrene beads are going to scatter a lot more light than a typical cell would.

According to studies published in Current Protocols In Cytometry, this situation can be even more severe for microvesicles. The study estimates an average and typical RI of a microvesicle to be approximately 1.39. Given the RI of a polystyrene bead at 1.59 and water at 1.333, a microvesicle’s scatter signal may be one to two orders of magnitude lower than that of the polystyrene bead. Here’s the bottom line—beads are not good at calibrating the scale of a scatter parameter in terms of particle size, regardless of whether the particle is a big one or a small one.

However, all is not lost when it comes to identifying microvesicles. Rather than use scatter as a trigger/threshold parameter to identify these kinds of particles and to measure them, the study suggests that fluorescence is a better choice. Small particles can be labeled with a universal dye that causes all membrane-bound particles in a suspension to fluoresce, allowing discrimination of microvesicles from other particles in the solution. There are nuances and caveats to this kind of labeling, but it can provide a much more robust way to identify membrane-bound microparticles than scatter alone.

Light scatter is a fundamental topic to flow cytometry and understanding how cell size and particle size affect light scatter is critical to performing proper flow cytometry experiments. By understanding how small particles affect forward scatter and side scatter, you can collect better data. Knowing what an obscuration bar is and what a refractive index is, as well as how refractive indices are affected by small particles, will help you design experiments that produce publishable data.

To learn more about getting your flow cytometry data published and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class
Tim Bushnell, PhD
Tim Bushnell, PhD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Similar Articles

Getting A New Flow Cytometer? Try Before You Buy (And 2 Other Tips)

Getting A New Flow Cytometer? Try Before You Buy (And 2 Other Tips)

By: Tim Bushnell, PhD

There are 4 major ways to sort cells. The first way can use magnetic beads coupled to antibodies and pass the cells through a magnetic field. The labeled cells will stick, and the unlabeled cells will remain in the supernatant. The second way is to use some sort of mechanical force like a flapper or air stream that separates the target cells from the bulk population. The third way is the recently introduced microfluidics sorter, which uses microfluidics channels to isolate the target cells. The last method, which is the most common––based on Fuwyler’s work––is the electrostatic cell sorter. This…

5-Point Guide To Buying A New Microscope For Your Lab

5-Point Guide To Buying A New Microscope For Your Lab

By: Heather Brown-Harding, PhD

Have you ever noticed how painful it can be to purchase a new microscope? It would be hard to miss – this can be a frustrating process. A lot of scientists and students consider the new microscope hunt quite scary for a variety of reasons. It might be that you’re worried you won’t get the right microscope and that you’ll regret it, or you may find that dealing with salespeople, in general, makes you kind of uncomfortable. But remember, salespeople are just human beings like you and me, and if we can treat them as such, the whole process of…

Ask These 7 Questions Before Purchasing A Flow Cytometer

Ask These 7 Questions Before Purchasing A Flow Cytometer

By: Tim Bushnell, PhD

I am still convinced that my first cell sorter was possessed. The number of issues that I had with the system remains hard for me to believe, even after all these years. It had been purchased, in part, from one vendor because the sales rep for a competitor was nowhere to be found. At that time, I admit I wasn’t overly diligent in my research process. Since then, I’ve pinpointed some critical questions that need to be answered before purchasing a new instrument. At the end of the process, a shiny new instrument will arrive at your facility. Make sure…

Instrument Quality Control For Reproducible Flow Cytometry Experiments

Instrument Quality Control For Reproducible Flow Cytometry Experiments

By: Tim Bushnell, PhD

The flow cytometer is an integral component of any flow cytometry experiment, and special attention should be paid to ensuring that it is working correctly and consistently. As an end-user, the researcher should be able to sit down at a machine and know that it is performing the same way today as it was yesterday and last week. Equally important is that if any changes in instrument performance have occured, the end-user knows how they have been addressed and corrected, rather than letting them fester and potentially affect the results. Quality control measurements can include a variety of targets, such…

How to Optimize Flow Cytometry Hardware For Rare Event Analysis

How to Optimize Flow Cytometry Hardware For Rare Event Analysis

By: Tim Bushnell, PhD

Preparing for rare event analysis requires an understanding of the power and limitation of the instrument to be used. From how fast to run the fluidics, to how the signal is processed to the number of gates that can be used in the sorting experiment, each factor impacts the outcome of the experiment.

3 Ways The ZE5 Cell Analyzer Accelerates Flow Cytometry Research Opportunities

3 Ways The ZE5 Cell Analyzer Accelerates Flow Cytometry Research Opportunities

By: Tim Bushnell, PhD

Some technological advances are incremental, while others are significant game-changing tools that offer the researcher the ability to significantly improve current assays while allowing for new and novel avenues of research to be performed. With speed, sensitivity, and capacity to spare, the ZE5 fits into the game-changing category. Reduced carryover, increased speed of acquisition, and a large number of parameters all open up new and novel assays while improving the quality and reproducibility of ongoing ones.

3 Advantages Of Using The ZE5 Cell Analyzer

3 Advantages Of Using The ZE5 Cell Analyzer

By: Tim Bushnell, PhD

Since the first laser was mounted to create the first flow cytometer, there has been a push for more - more lasers, more detectors, more colors. As a result, today’s researchers require a large number of lasers and detectors to ensure current panels can be run and new, expanded panels can be developed. This can be problematic because, in general, making one decision to improve a cell analyzer can limit the analyzer in other ways. It may seem like an impossible task, but the team of Bio-Rad and Propel Laboratories, collaborated to bring the ZE5™ Cell Analyzer to the market…

3 Advantages FCS Express 6 Has Over Other Flow Cytometry Data Analysis Software Programs

3 Advantages FCS Express 6 Has Over Other Flow Cytometry Data Analysis Software Programs

By: Tim Bushnell, PhD

FCS Express is the ideal data analysis software program to use when analyzing your flow cytometry experiments because it is the most user-friendly program available that is both aligned with current data analysis best practices and maintains rigorous quality control standards.

How To Use A Threshold To Reduce Background Noise In Flow Cytometry

How To Use A Threshold To Reduce Background Noise In Flow Cytometry

By: Tim Bushnell, PhD

Getting a clear signal with reduced noise is an essential component to good data. Adding a threshold when acquiring flow cytometry data is one way to do that. It reduces the number of events by setting a bar that a signal pulse must clear before it is counted as an event. Depending on the importance of the data, the downstream applications for the data (or sorted cells) will dictate how critical the threshold is. In combination with proper sample preparation, appropriate thresholding will reduce debris and ensure best outcome.

Top Technical Training eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.