How To Set And Monitor Optimal Voltages For A Flow Cytometry Experiment

When a researcher first sits down in front of the flow cytometer, they are faced with several choices.

One of the most confusing is, “What voltage do I set my detector at?”

Unless the researcher is running a fixed voltage system (Accuri, and others), this choice can dramatically impact the sensitivity of the instrument, making or breaking the experiment.

In the days of the analog flow cytometers, voltages were set by placing a quadrant gate on a bivariant plot, with the lower left quadrant encompassing the first log decade. Unstained cells would be run on the instrument and the PMT voltage set until these cells were contained within that lower left quadrant.

Flow Cytometry voltage settings using old school methods

Figure 1: Schematic of setting voltages ‘Old School’.

Because of how the data was processed in older generation instruments, a portion of the population was ‘off-scale’ and accumulated in the first channel.

Using this method, the highly autofluorescent cells would drive the voltage, often causing a compression of the less autofluorescent cells on the axis. However, this was not always obvious because of how the data was plotted.

It should be remembered that with these systems, the data was log transformed (and compensated) in the hardware, and stored that way. This prevented much post-acquisition manipulation.

Enter the digital cytometer age, in which the data is processed post-acquisition.

This paradigm change in data acquisition and storage also meant that traditional methods for setting voltages needed to be reviewed and new methods developed.

1. Peak 2 optimization.

All PMTs have a sensitive sweet spot: a voltage at which the detector is most sensitive for the fluorochrome emission being measured.

The first paper to discuss a practical application of determining the voltage sweet spot was published by Maecker and Trotter in 2006. In this paper, the authors discuss what is termed the ‘Peak 2’ method for determining sensitivity.

In this method, a very dim particle (peak 2 of the Spherotech 8 peak bead set) is run over a series of voltages, and the Coefficient of Variance (CV) of the bead is plotted versus the voltage.

A typical graph is shown in Figure 2.

Peak 2 voltage setting in a flow cytometer

Figure 2: Results of a Peak 2 optimization of a PMT.

As PMT voltage increases, CV of the beads decreases until it hits an inflection point, and there is no improvement in the CVs from that point. The red arrow indicates the ‘optimal’ sensitivity.

This graph is interpreted by identifying the inflection point on the graph (shown in the red arrow).

Increasing the voltage from this point does not improve PMT sensitivity (with a caveat, discussed later), so the best voltage to start at is just below that inflection point.

2. Cytometry setup and tracking.

BD ran with this idea, and implemented a different method in their Cytometry Setup and Tracking.

When a CS&T baseline is run, the optimal voltage is determined by first determining the Standard Deviation of Electronic Noise (SDEN).

The software calculates the voltage necessary to set the baseline at 10X SDEN. The results of this method are stored in the CS&T report, and a typical curve is shown in Figure 3.

CST flow cytometer optimization

Figure 3: Results of a CS&T optimization baseline report.

If you have a digital machine, made by anyone other than BD, what can you do?

Option one is to follow the peak 2 method, which works very well.

A second option is to measure the electronic noise in your system and generate a voltage via CS&T.

One method to measure electronic noise is to run a negative particle, (better yet, unstained cells), over a voltage range. Then, generate a plot of (1/mean2 vs CV2), and calculate a regression line.

The slope of the line is the Variance of Electronic Noise, and the standard deviation is the square root of the variance. This gives you a channel value for electronic noise. Multiply by ten to get the target value for each channel.

Now put your unstained cells on the instrument, adjust to the target value, and away you go.

Results of calculating the SDEN using PBMCs run over a voltage series

Figure 4: Calculating SDEN. Results of calculating the SDEN using PBMCs run over a voltage series.

You can improve these target values by performing a voltage optimization using the target cells stained with the properly titered reagents for the specific panel.

This can help improve sensitivity because of the fact that the spectrum of fluorochrome of interest in your panel may not be accurately modeled using the beads.

As shown in the figure below, two different fluorochromes give very different results.

PMT optimization for correct flow cytometry operation

Figure 5: PMT optimization.

Cells were labeled with optimal concentrations of antibodies of either CY7-APC (Left) or BV650 (Right), and a voltage optimization was performed, starting below the recommended peak 2 value.

The separation index was plotted against the voltage and the curves above generated.

Particularly with the newer dyes, an improvement in the separation index identifies a better voltage for this cell/fluorochrome combination.

3. Setting consistent target voltages.

The last step in voltage setting on a digital instrument is to have some way to consistently set those target voltages over time, without having to go through a long, tedious process.

Beads again come to the rescue. In this case, a very bright bead (some favor the 6th peak of the 8 peak bead-set, but you can choose your favorite), that is fluorescent in the channels to be used is run when the optimal voltages are identified. This results in a value for each detector — the ‘target value’.

Then set up a template with the appropriate plots and target values and remember to save it for later use!

When you come back to the instrument, open up the template and run this bright bead. Adjust the voltages (as necessary) to achieve the target values in each channel and you’re good to start acquiring your samples.

The other advantage of this method is that you can become cross-platform compatible.

The best way to take out the fear and agony of setting voltages is to use some optimization methods. The peak 2 method, as described above, is a useful and robust method of identifying optimal PMT voltage ranges. Refining that to the voltage walk with the actual cells and fluorochromes of interest will further improve sensitivity, which is especially critical for rare cell populations or emergent antigens (like activation markers). Of course, don’t forget to set-up a way to monitor and maintain those voltage settings.

To learn more about how to set and monitor optimal voltages for a flow cytometry experiment, and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class
Tim Bushnell, PhD
Tim Bushnell, PhD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Similar Articles

Which Fluorophores To Use For Your Microscopy Experiment

Which Fluorophores To Use For Your Microscopy Experiment

By: Heather Brown-Harding, PhD

Fluorophore selection is important. I have often been asked by my facility users which fluorophore is best suited for their experiments. The answer to this is mostly dependent on whether they are using a widefield microscope with set excitation/emission cubes or a laser based system that lets you select the laser and the emission window. Once you have narrowed down which fluorophores you can excite and collect the correct emission, you can further refine the specific fluorophore that is best for your experiment.  In this blog  we will discuss how to determine what can work with your microscope, and how…

How To Optimize Instrument Voltage For Flow Cytometry Experiments  (Part 3 Of 6)

How To Optimize Instrument Voltage For Flow Cytometry Experiments (Part 3 Of 6)

By: Tim Bushnell, PhD

As we continue to explore the steps involved in optimizing a flow cytometry experiment, we turn our attention to the detectors and optimizing sensitivity: instrument voltage optimization.  This is important as we want to ensure that we can make as sensitive a measurement as possible.  This requires us to know the optimal sensitivity of our instrument, and how our stained cells are resolved based on that voltage.  Let’s start by asking the question what makes a good voltage?  Joe Trotter, from the BD Biosciences Advanced Technology Group, once suggested the following:  Electronic noise effects resolution sensitivity   A good minimal PMT…

How To Profile DNA And RNA Expression Using Next Generation Sequencing (Part-2)

How To Profile DNA And RNA Expression Using Next Generation Sequencing (Part-2)

By: Deepak Kumar, PhD

In the first blog of this series, we explored the power of sequencing the genome at various levels. We also dealt with how the characterization of the RNA expression levels helps us to understand the changes at the genome level. These changes impact the downstream expression of the target genes. In this blog, we will explore how NGS sequencing can help us comprehend DNA modification that affect the expression pattern of the given genes (epigenetic profiling) as well as characterizing the DNA-protein interactions that allow for the identification of genes that may be regulated by a given protein.  DNA Methylation Profiling…

4 No Cost Ways To Improve Your Microscopy Image Quality

4 No Cost Ways To Improve Your Microscopy Image Quality

By: Heather Brown-Harding, PhD

Image quality is critical for accurate and reproducible data. Many people get stuck on the magnification of the objective or on using a confocal instead of a widefield microscope. There are several other factors that affect the image quality such as the numerical aperture of the objective, the signal-to-noise ratio of the system, or the brightness of the sample.  Numerical aperture is the ability of an objective to collect light from a sample, but it contributes to two key formulas that will affect your image quality. The first is the theoretical resolution of the objective. It is expressed with the…

How To Profile DNA And RNA Expression Using Next Generation Sequencing

How To Profile DNA And RNA Expression Using Next Generation Sequencing

By: Deepak Kumar, PhD

Why is Next Generation Sequencing so powerful to explore and answer both clinical and research questions. With the ability to sequence whole genomes, identifying novel changes between individuals, to exploring what RNA sequences are being expressed, or to examine DNA modifications and protein-DNA interactions occurring that can help researchers better understand the complex regulation of transcription. This, in turn, allows them to characterize changes during different disease states, which can suggest a way to treat said disease.  Over the next two blogs, I will highlight these different methods along with illustrating how these can help clinical diagnostics as well as…

Optimizing Flow Cytometry Experiments - Part 2         How To Block Samples (Sample Blocking)

Optimizing Flow Cytometry Experiments - Part 2 How To Block Samples (Sample Blocking)

By: Tim Bushnell, PhD

In my previous blog on  experimental optimization, we discussed the idea of identifying the best antibody concentration for staining the cells. We did this through a process called titration, which  focuses on finding the best signal-to-noise ratio at the lowest antibody concentration. In this blog we will deal with sample blocking As a reminder, there are two other major binding concerns with antibodies. The first is the specific binding of the Fc fragment of the antibody to the Fc Receptor expressed on some cells. This protein is critical for the process of destroying microbes or other cells that have been…

What Is Total Internal Reflection Fluorescence (TIRF) Microscopy & Is It Right For You?

What Is Total Internal Reflection Fluorescence (TIRF) Microscopy & Is It Right For You?

By: Heather Brown-Harding, PhD

TIRF is not as common as other microscopy based techniques due to certain restrictions. We will discuss these restrictions, then analyze why it might be perfect for your experiment.  TIRF relies on an evanescent wave, created through a critical angle of coherent light (i.e. laser) that reaches a refractive index mismatch.  What does it mean in practice?  A high angle laser reflects off the interface of the coverslip and the sample. Although the depth that this wave penetrates is dependent on the wavelength of the light, in practice it is approximately 50-300nm from the coverslip. Therefore, the cell membrane is…

What Is Next Generation Sequencing (NGS) And How Is It Used In Drug Development

What Is Next Generation Sequencing (NGS) And How Is It Used In Drug Development

By: Deepak Kumar, PhD

NGS methodologies have been used to produce high-throughput sequence data. These data with appropriate computational analyses facilitate variant identification and prove to be extremely valuable in pharmaceutical industries and clinical practice for developing drug molecules inhibiting disease progression. Thus, by providing a comprehensive profile of an individual’s variome — particularly that of clinical relevance consisting of pathogenic variants — NGS helps in determining new disease genes. The information thus obtained on genetic variations and the target disease genes can be used by the Pharma companies to develop drugs impeding these variants and their disease-causing effect. However simple this may allude…

How To Determine The Optimal Antibody Concentration For Your Flow Cytometry Experiment (Part 1 of 6)

How To Determine The Optimal Antibody Concentration For Your Flow Cytometry Experiment (Part 1 of 6)

By: Tim Bushnell, PhD

Over the next series of blog posts, we will explore the different aspects of optimizing a polychromatic flow cytometry panel. These steps range from figuring out the best voltage to use, which controls are critical for data interpretation, what quality control tools can be integrated into the assay; how to block cells, and more. This blog will focus on determining the optimal antibody concentration.  As a reminder about the antibody structure, a schematic of an antibody is shown below.  Figure 1: Schematic of an antibody. Figure from Wikipedia. The antibody is composed of two heavy chains and two light chains that…

Top Technical Training eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.