3 Types Of Flow Cytometry Beads That Will Help Get Your Data Published

When starting a flow cytometry experiment, you should begin with the end in mind.

What data do you need to generate to support your other data? What data would really impress the reviewers in your study section and get your lab the grant you need? What data do you need to land your first tenure-track position?

Planning with your end goal in mind is important, but don’t forget the details along the way. Your cells are important, but ensuring your experiment is well set-up and your instrument properly maintained is just as important as elegant experimental design and good writing.

The 3 Best Flow Cytometry Beads

To make certain your instrument is set up correctly for your experiments, manufacturers have developed defined polystyrene beads.

These beads’ consistent nature helps you to assess how your instrument is behaving, helps you set up proper compensation matrices, and helps you generate volumetric counts of your cell populations. The best beads of the bunch in each of these 3 areas are highlighted below…

1. Quality control beads. 

Manufacturers make certain that when your instrument is built and installed that the optical alignment is ideal. They do this with beads with multiple sizes and/or fluorescence levels, exactly like the ones you should be using. After installation and between preventative maintenance appointments, it is up to you (or your core staff) to track the performance of your instrument.

Alignment, sensitivity, and fluidic quality control beads should be run at least daily to ensure that with the same wattage on the laser and the same voltage applied to the detector you return the same median fluorescence (MFI).

Most new cytometers have an automated system to perform this activity, but even if yours does not, it is quite easy to generate your own Levy-Jennings plots.

There are several things to look out for when running your quality control beads. One thing to look out for is any change in the amount of voltage that needs to be applied to get the same MFI reading. Such a reading could indicate a loss of laser power if it occurs across all detectors for a given laser, or it could indicate a failing detector if the reading is isolated to a single PMT.

In addition, look out for changes in the coefficient of variation (CV) of your bead populations. Increased CV means less sensitivity and could point to misalignment of the laser, or issues with detection optics.

Our favorite quality control beads are BD’s CS&T beads and Spherotech’s multiple peak beads. The BD beads are really designed for their instruments using DiVa software, but they can certainly be used on other systems.

2. Compensation “capture” beads. 

Compensation was a lot easier when you only had to work with FITC and PE. All you had to do was get some beads impregnated with those compounds, set your voltages, adjust your compensation matrix to correct for overlap, and voilà —you’re done.

Now, even beginners want to do 5+ colors. This makes antibody-binding compensation beads a necessity. 

Beads with fluorochrome impregnated in the bead are still available, but compensation “capture” beads are ideal. The best of these beads bind antibodies of multiple isotypes from multiple species and give you a very bright positive signal from which to calculate your compensation matrix. These compensation beads also allow you to use the same fluorochromes you’re using in your experiment.

Using compensation capture beads is especially important when using tandem dyes as lot-to-lot variability and time-dependent changes in storage can change the spectral properties of these dyes.

“But my cells are free and work just fine!”

Sorry Charlie, you’re wrong.

One of the most important rules of good compensation is you want your compensation controls to be as bright, or brighter, than your actual experimental samples.

Beads will almost always be brighter (and therefore more accurate) in calculating a compensation matrix.

Another rule of compensation is that the colors have to be identical. If you’re using FITC in your experimental sample then you must use FITC in your compensation control. If you’re using tandem-dyes in your sample, you MUST use the same tandem dye in your controls, and so on. Using antibody capture beads ensures you’re using the same fluorochrome as is in your sample, ensuring compensation success.

One final note on compensation beads—compensation is the property of the fluorochrome, not the antigen.  

Many novice flow cytometrists will make the mistake of staining the compensation beads with the same amount of antibody as is on their cells. This can lead to disaster, especially when the compensation control is off-scale on the instrument. If this happens, avoid the temptation to decrease the PMT voltage as this will just reduce the sensitivity on your cells (where it really matters).

The real solution is to reduce the amount of fluorochrome on the beads. We recommend you use 1/10 the amount of antibody on the beads as you do on the cells.  Of course, this may vary depending on the antibody, and you must still make sure that the compensation signal is as bright as the signal on your cells.

Once again, if your compensation control is off-scale—DO NOT TURN THE VOLTAGE DOWN. Instead, re-stain the beads with less antigen.  

Our favorite compensation beads are eBioscience’s Ultracomp beads. A lot of companies make capture beads, but these are the only ones I have seen that have such a wide range of species and isotype binding. Otherwise, be very careful about the species binding characteristics of what you buy.

3. Counting beads.

Counting beads are beads that come in multiple sizes, with or without fluorescence. The key is that these beads are shipped at a defined concentration. This allows you to set your stop gate to acquire a set number of events and use the events to calculate back to your cell concentration in the original sample.

Our favorite counting beads are Life Technologies Accucheck beads. These and other counting beads come in fluorescent and nonfluorescent forms but the fluorescent ones tend to be best. The fluorescence makes them very easy to gate for compared to gating on their scatter signal alone.

As seen in the example below, you could have a hard time gating on your count beads by scatter alone, as they are very close to your cells.

counting beads for flow cytometry | Expert Cytometry | flow cytometry calibration beads

It can be hard enough to design your experiment with the proper controls, harvest your samples and develop your protocols. Make sure that you have an instrument set up that is stable, a good compensation matrix, and if you are at all interested in true counts of cells, the beads to make that accurate. If you are in a core environment, odds are very good that they are already running quality control daily on all instruments, and many people are not concerned with volumetric counts, but if you take nothing else from this article, make sure you are using capture beads to generate your compensation matrix.

Bonus—PMT Voltage Optimization Beads

Another one of our favorite beads are the Spherotech peak-2 bead (that is the second bead of the 8 peak bead-set—RCP-30-5A-2). These beads are used to optimize the PMT voltage settings on a flow cytometer based on this paper by Maecker and Trotter (2006).

To use these beads, simply run a voltage series on each PMT (typically 300-900 volts) and plot the CV of the data versus the voltage range. As a result, you’ll generate data similar to the data below.

bd compensation beads | Expert Cytometry | flow cytometry compensation beads

As voltage is increased, the CV of the beads (the measure of the spread of the data) decreases to an inflection point. At this point, the CV no longer increases, indicating that increasing voltages leads to increases in sensitivity of the detector.

The above flow cytometry beads will help ensure your experiment is well set-up and your instrument properly maintained. Alignment, sensitivity, and fluidic quality control beads will help you to ensure that with the same wattage on the laser and the same voltage applied to the detector returns the same MFI. The right compensation capture beads will bind antibodies of multiple isotypes from multiple species and give you a very bright positive signal from which you can calculate a correct compensation matrix. Finally, the use of counting beads allows you to easily calculate your cell concentration in your sample. Together, these flow cytometry beads will make your life easier and help you get your data published.

To learn more about the right flow cytometry reagents to use for your experiments and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class
Tim Bushnell, PhD
Tim Bushnell, PhD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Similar Articles

Getting A New Flow Cytometer? Try Before You Buy (And 2 Other Tips)

Getting A New Flow Cytometer? Try Before You Buy (And 2 Other Tips)

By: Tim Bushnell, PhD

There are 4 major ways to sort cells. The first way can use magnetic beads coupled to antibodies and pass the cells through a magnetic field. The labeled cells will stick, and the unlabeled cells will remain in the supernatant. The second way is to use some sort of mechanical force like a flapper or air stream that separates the target cells from the bulk population. The third way is the recently introduced microfluidics sorter, which uses microfluidics channels to isolate the target cells. The last method, which is the most common––based on Fuwyler’s work––is the electrostatic cell sorter. This…

5-Point Guide To Buying A New Microscope For Your Lab

5-Point Guide To Buying A New Microscope For Your Lab

By: Heather Brown-Harding, PhD

Have you ever noticed how painful it can be to purchase a new microscope? It would be hard to miss – this can be a frustrating process. A lot of scientists and students consider the new microscope hunt quite scary for a variety of reasons. It might be that you’re worried you won’t get the right microscope and that you’ll regret it, or you may find that dealing with salespeople, in general, makes you kind of uncomfortable. But remember, salespeople are just human beings like you and me, and if we can treat them as such, the whole process of…

Ask These 7 Questions Before Purchasing A Flow Cytometer

Ask These 7 Questions Before Purchasing A Flow Cytometer

By: Tim Bushnell, PhD

I am still convinced that my first cell sorter was possessed. The number of issues that I had with the system remains hard for me to believe, even after all these years. It had been purchased, in part, from one vendor because the sales rep for a competitor was nowhere to be found. At that time, I admit I wasn’t overly diligent in my research process. Since then, I’ve pinpointed some critical questions that need to be answered before purchasing a new instrument. At the end of the process, a shiny new instrument will arrive at your facility. Make sure…

Instrument Quality Control For Reproducible Flow Cytometry Experiments

Instrument Quality Control For Reproducible Flow Cytometry Experiments

By: Tim Bushnell, PhD

The flow cytometer is an integral component of any flow cytometry experiment, and special attention should be paid to ensuring that it is working correctly and consistently. As an end-user, the researcher should be able to sit down at a machine and know that it is performing the same way today as it was yesterday and last week. Equally important is that if any changes in instrument performance have occured, the end-user knows how they have been addressed and corrected, rather than letting them fester and potentially affect the results. Quality control measurements can include a variety of targets, such…

How to Optimize Flow Cytometry Hardware For Rare Event Analysis

How to Optimize Flow Cytometry Hardware For Rare Event Analysis

By: Tim Bushnell, PhD

Preparing for rare event analysis requires an understanding of the power and limitation of the instrument to be used. From how fast to run the fluidics, to how the signal is processed to the number of gates that can be used in the sorting experiment, each factor impacts the outcome of the experiment.

3 Ways The ZE5 Cell Analyzer Accelerates Flow Cytometry Research Opportunities

3 Ways The ZE5 Cell Analyzer Accelerates Flow Cytometry Research Opportunities

By: Tim Bushnell, PhD

Some technological advances are incremental, while others are significant game-changing tools that offer the researcher the ability to significantly improve current assays while allowing for new and novel avenues of research to be performed. With speed, sensitivity, and capacity to spare, the ZE5 fits into the game-changing category. Reduced carryover, increased speed of acquisition, and a large number of parameters all open up new and novel assays while improving the quality and reproducibility of ongoing ones.

3 Advantages Of Using The ZE5 Cell Analyzer

3 Advantages Of Using The ZE5 Cell Analyzer

By: Tim Bushnell, PhD

Since the first laser was mounted to create the first flow cytometer, there has been a push for more - more lasers, more detectors, more colors. As a result, today’s researchers require a large number of lasers and detectors to ensure current panels can be run and new, expanded panels can be developed. This can be problematic because, in general, making one decision to improve a cell analyzer can limit the analyzer in other ways. It may seem like an impossible task, but the team of Bio-Rad and Propel Laboratories, collaborated to bring the ZE5™ Cell Analyzer to the market…

3 Advantages FCS Express 6 Has Over Other Flow Cytometry Data Analysis Software Programs

3 Advantages FCS Express 6 Has Over Other Flow Cytometry Data Analysis Software Programs

By: Tim Bushnell, PhD

FCS Express is the ideal data analysis software program to use when analyzing your flow cytometry experiments because it is the most user-friendly program available that is both aligned with current data analysis best practices and maintains rigorous quality control standards.

How To Use A Threshold To Reduce Background Noise In Flow Cytometry

How To Use A Threshold To Reduce Background Noise In Flow Cytometry

By: Tim Bushnell, PhD

Getting a clear signal with reduced noise is an essential component to good data. Adding a threshold when acquiring flow cytometry data is one way to do that. It reduces the number of events by setting a bar that a signal pulse must clear before it is counted as an event. Depending on the importance of the data, the downstream applications for the data (or sorted cells) will dictate how critical the threshold is. In combination with proper sample preparation, appropriate thresholding will reduce debris and ensure best outcome.

Top Technical Training eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.