Avoid Flow Cytometry Faux Pas: How To Set Voltage The Right Way

The Shift from Analog To Digital Cytometer Age

Analog instruments processed data differently than the current generation of digital instruments. With analog systems, if the populations were “off-scale,” especially at the low end of the scale, the data accumulated in the first channel. When setting voltage, highly autofluorescent cells would drive the voltage, and it was not uncommon for less autofluorescent cells on the axis to get compressed. Unfortunately, due to the way the data were plotted, this side effect was not always clear to observers.

With these older systems, the data were log-transformed and compensated in the hardware preventing much manipulation post-acquisition. Fortunately, with the advent of the digital cytometer age, data could finally be processed post-acquisition. This monumental shift meant it was time to review traditional voltage-setting methods, and new methods would need to be developed.

Methods To Set Voltage

There are a few different ways to set voltage. There is the historical method of putting the negatives in the first decade; you can calculate the standard deviation of electronic noise, set your negatives at 2.5-to-10x that value. You can use the OEM method to set voltages. You can use the Peak 2 method. Or you can use the “2x brighter” method. Last but not least, you can measure Q and B to help you define your values. We generally do not recommend the “SWG” method (scientific wild guess). 

Of course, at the end of the day, the important thing about setting voltages is that you identify a method and stick with it. So what makes a good voltage? That’s the question I’m going to address in this article. For one, the ideal voltage will give the best signal-to-noise ratio for a given fluorochrome. It’s going to make sure that the fluorochrome is within the linear range of the detector, and it will ensure that the fluorochrome is not off-scale.

Voltage Optimization: The Peak 2 Method

My preferred method for setting voltages involves a two-fold process. First, you need to characterize the PMTs. To do this, use the Peak 2 method from Maecker and Trotter’s paper. For this method, you run a dimly fluorescent particle (such as RCP-30-5A-2 from Spherotech) over a voltage range. Next, you calculate the rCV at each point, then plot and identify the inflection point – that’s your minimum starting value.

Figure 1 shows the results of the Peak 2 method. Beginning at a point of low voltage and high CV, it progressed to an inflection, marking the starting point of the optimal voltage range. Notice that there is a change in the slope of the curve. Above this point, you’ll also notice that the signal doesn’t improve – you don’t see a change. What you have to do is use the inflection point as your minimal operating voltage.

Figure 1: Determining voltage settings using the Peak 2 method.

These voltages can be refined by determining experimental specific voltages, which reflect the actual fluorochromes and cells that are being used in the experiment. To replicate this process, run the cells using the titer antibody concentration. Then run the voltage series, starting from this minimal starting voltage, and calculate the staining index. Basically, you’re looking at the ratio between the negative and the positive. You can do this using a foldover background, a staining index, or a separation index. I have never actually noticed a difference between any of the 3 methods, so I stick to using the staining index.

The next step is to plot the staining index vs. the voltage and identify your optimal voltage, as shown in figure 2. As you will notice, not every channel will be improved. 

Figure 2: Voltage optimization of two different fluorochromes. 

Cells are stained at the optimal titration concentration and run across a voltage range. Next, plot the staining index against the voltages. In figure 2, the peak 2 value is marked with a blue arrow. The optimal voltage is circled in green. On the left figure, as the voltage is increased, the signal first goes out of the linear range of the detector, which means we can’t use the data as we cannot properly compensate, thus, the peak 2 voltage is optimal. The figure on the right, on the other hand, increasing the voltage identifies a voltage where we get a 20% increase in the SI. To the right, the SI decreases because of an increase in the background signal. 

Tracking Voltages Over Time

Finally, you need to develop a way to consistently set target voltages over time (without having to go through a long, tedious process). After the optimal voltages are identified, run a bright, fluorescent bead such as the Spherotech 6th peak of the 8 peak headset. This will give you a “target value”, which is used when you set up a template with the appropriate plots and target values. The next time you use the instrument, open your template, and run that bright bead, adjusting the voltages to achieve the target values in each channel. Once that is done, you will be ready to start acquiring your samples. An example template is shown below.

Figure 3: Example template for instrument setup and tracking voltages over time. 

This method has another advantage: you can become cross-platform compatible. The best way to simplify the voltage-setting process (and eliminate accompanying dread) is to use some optimization methods. 

For those using the CytoFLEX™, a variation of this method has been recently published to aid in optimizing this platform. 

Concluding Remarks

The Peak 2 method—described earlier—is a useful and robust method by which you can identify optimal PMT voltage ranges. You can even further improve sensitivity by refining that to the voltage walk with the actual cells and fluorochromes of interest. This is especially critical for rare cell populations—or emergent antigens like activation markers. And to conclude, make sure you set up a way to monitor and maintain your voltage settings.

To learn more about important control measures for your flow cytometry lab, and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class
Tim Bushnell, PhD
Tim Bushnell, PhD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Similar Articles

How To Profile DNA And RNA Expression Using Next Generation Sequencing

How To Profile DNA And RNA Expression Using Next Generation Sequencing

By: Deepak Kumar, PhD

Why is Next Generation Sequencing so powerful to explore and answer both clinical and research questions. With the ability to sequence whole genomes, identifying novel changes between individuals, to exploring what RNA sequences are being expressed, or to examine DNA modifications and protein-DNA interactions occurring that can help researchers better understand the complex regulation of transcription. This, in turn, allows them to characterize changes during different disease states, which can suggest a way to treat said disease.  Over the next two blogs, I will highlight these different methods along with illustrating how these can help clinical diagnostics as well as…

Optimizing Flow Cytometry Experiments - Part 2         How To Block Samples (Sample Blocking)

Optimizing Flow Cytometry Experiments - Part 2 How To Block Samples (Sample Blocking)

By: Tim Bushnell, PhD

In my previous blog on  experimental optimization, we discussed the idea of identifying the best antibody concentration for staining the cells. We did this through a process called titration, which  focuses on finding the best signal-to-noise ratio at the lowest antibody concentration. In this blog we will deal with sample blocking As a reminder, there are two other major binding concerns with antibodies. The first is the specific binding of the Fc fragment of the antibody to the Fc Receptor expressed on some cells. This protein is critical for the process of destroying microbes or other cells that have been…

What Is Next Generation Sequencing (NGS) And How Is It Used In Drug Development

What Is Next Generation Sequencing (NGS) And How Is It Used In Drug Development

By: Deepak Kumar, PhD

NGS methodologies have been used to produce high-throughput sequence data. These data with appropriate computational analyses facilitate variant identification and prove to be extremely valuable in pharmaceutical industries and clinical practice for developing drug molecules inhibiting disease progression. Thus, by providing a comprehensive profile of an individual’s variome — particularly that of clinical relevance consisting of pathogenic variants — NGS helps in determining new disease genes. The information thus obtained on genetic variations and the target disease genes can be used by the Pharma companies to develop drugs impeding these variants and their disease-causing effect. However simple this may allude…

How To Determine The Optimal Antibody Concentration For Your Flow Cytometry Experiment (Part 1 of 6)

How To Determine The Optimal Antibody Concentration For Your Flow Cytometry Experiment (Part 1 of 6)

By: Tim Bushnell, PhD

Over the next series of blog posts, we will explore the different aspects of optimizing a polychromatic flow cytometry panel. These steps range from figuring out the best voltage to use, which controls are critical for data interpretation, what quality control tools can be integrated into the assay; how to block cells, and more. This blog will focus on determining the optimal antibody concentration.  As a reminder about the antibody structure, a schematic of an antibody is shown below.  Figure 1: Schematic of an antibody. Figure from Wikipedia. The antibody is composed of two heavy chains and two light chains that…

2020 - A Year Turned Upside Down

2020 - A Year Turned Upside Down

By: Tim Bushnell, PhD

What an incredible year 2020 has been. It started off like any other year and bam SARS-CoV-2 (aka COVID 19) entered the equation, bringing chaos and havoc to the world. Things kept changing overnight as new rules and regulations popped up. Masking, quarantine, and flatten the curve became common words in the news. How we met, how we interacted changed almost overnight. Throughout all of this, as we look to 2021, there is hope and optimism. Multiple vaccines have been developed, building on years of research into the SARS-CoV virus, with some approved for human use, and others on the horizon.…

Brightness Is In The Eye Of The Detector - What To Consider When Designing Your Panel

Brightness Is In The Eye Of The Detector - What To Consider When Designing Your Panel

By: Tim Bushnell, PhD

The heart and soul of the flow cytometry experiment is the ‘panel.’ The unique combinations of antibodies, antigens, fluorochromes, and other reagents are central to identifying the cells of interest and extracting the data necessary to answer the question at hand. Designing the right panel for flow cytometry is essential for detecting different modalities. The more parameters that can be interrogated will yield more information about the target cells. Current instruments can measure as many as 40 different parameters simultaneously. This is exciting, as it allows for more complex questions to be studied. Panel design is also valuable for precious samples,…

Tools to Improve Your Panel Design – Determining Antigen Density

Tools to Improve Your Panel Design – Determining Antigen Density

By: Tim Bushnell, PhD

When a researcher chooses to use flow cytometry to answer a scientific question, they first have to build a polychromatic panel that will take advantage of the power of the technology and experimental design. When we set up to use flow cytometry to answer a scientific question, we have to design a polychromatic panel that will allow us to identify the cells of interest – the target of the research.  To identify these cells, we need to build a panel that takes advantage of the relative brightness of the fluorochromes, the expression level of the different proteins on the cell,…

The 5 Fundamental Methods For Imaging Nucleic Acids

The 5 Fundamental Methods For Imaging Nucleic Acids

By: Heather Brown-Harding, PhD

There are 4 major ways to sort cells. The first way can use magnetic beads coupled to antibodies and pass the cells through a magnetic field. The labeled cells will stick, and the unlabeled cells will remain in the supernatant. The second way is to use some sort of mechanical force like a flapper or air stream that separates the target cells from the bulk population. The third way is the recently introduced microfluidics sorter, which uses microfluidics channels to isolate the target cells. The last method, which is the most common––based on Fuwyler’s work––is the electrostatic cell sorter. This…

This Is How Full Spectrum Cytometry Works

This Is How Full Spectrum Cytometry Works

By: Tim Bushnell, PhD

There are 4 major ways to sort cells. The first way can use magnetic beads coupled to antibodies and pass the cells through a magnetic field. The labeled cells will stick, and the unlabeled cells will remain in the supernatant. The second way is to use some sort of mechanical force like a flapper or air stream that separates the target cells from the bulk population. The third way is the recently introduced microfluidics sorter, which uses microfluidics channels to isolate the target cells. The last method, which is the most common––based on Fuwyler’s work––is the electrostatic cell sorter. This…

Top Technical Training eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.