1 – Instrumentation

Cell Sorting

By: Tim Bushnell, PhD

Cell Sorting is the process of isolating cells after the identification of the cells using the principles of flow cytometry. The upstream components of the cell sorter are common to all flow cytometers. The difference comes in what is done with the cells after they have been interrogated and identified. The stream is vibrated to generate thousands of individual droplets (as many as 90,000 or more), a fraction of which contain a cell. Those droplets that contain a cell of interest can beelectrically charged, as as the pass into an electric field, are deflected to the final receptacle, as shown…

Top Cell Sorter

By: Tim Bushnell, PhD

The question always arises as to what is the top cell sorter on the market. This question is a difficult one to generalize because there are several considerations that need to be made in choosing a cell sorter. What are the sorting needs of the investigators? If all the investigators do is sort GFP positive cells, then a simple sorter like the Bio-Rad S3 fits the bill. On the other hand, of the investigators need to do 4-12 color experiments with four way sorting, the choice becomes more muddled. Who will operate the instrument? Is the instrument going to be…

Yellow Laser

By: Tim Bushnell, PhD

A laser type in a flow cytometer with a wavelength of about 560nm. The green and yellow laser are more effective at exciting PE and its tandems than the traditional blue laser. The yellow laser is also often used to excite the “fruit” dyes like mCherry. For more information, please review this journal article: Telford  W,  Murga  M,  Hawley  T, et.al. (2005). DPSS yellow-green 561-nm lasers for improved fluorochrome detection by flow cytometry. Cytometry. 68A: 36-44

Green Laser

By: Tim Bushnell, PhD

The laser type in flow cytometers with a wavelength of around 530nm. Standard “green” lasers are about 532nm, but vary between 530nm and 535nm usually. The green and yellow laser are more effective at exciting PE and its tandems than the traditional blue laser.

UV Laser

By: Tim Bushnell, PhD

A laser with a wavelength in the UV range. Typically in flow cytometers, the UV laser has a wavelength of 350nm or 355nm. Some have a wavelength of 375nm.

Violet Laser

By: Tim Bushnell, PhD

Another very common laser after the “blue” and “red” laser in flow cytometers. A “violet” laser in flow cytometry typically is referred to as the 405 because most flow cytometers use a violet laser with a wavelength of 405nm. Pacific Blue and Pacific Orange are the most common fluorophores used with this laser, but Brilliant Violet fluors are gaining popularity.

Red Laser

By: Tim Bushnell, PhD

The second most common laser in a flow cytometer after the “blue” laser. The “red” laser typically has a wavelength of 633nm, but new flow cytometers are starting to use a “red” laser with a wavelength of 640nm. The most common fluorophores excited and detected off this laser are APC, Alexa Fluor 660, Alexa Fluor 700, and APC-tandems.

Blue Laser

By: Tim Bushnell, PhD

The most common laser type in a flow cytometer. Typically, this laser has a wavelength of 488nm in flow cytometers.  In fact, the term “Blue” laser is often interchanged with “488” laser. Frequently used fluorophores excited and detected by this laser are FITC, Alexa Fluor 488, PE, PerCP, and their tandems.

Bandpass Filter

By: Tim Bushnell, PhD

A filter that allows light between a set wavelength to pass through and reflects light above and below the set wavelength. For example, a bandpass filter with a wavelength of 550/40nm would allow light between 530nm and 570nm to pass through, but reflect light below 530nm and above 570nm.

Shortpass Filter

By: Tim Bushnell, PhD

A filter that allows light over a set wavelength to pass through and reflects light above the set wavelength. For example, a shortpass filter with a wavelength of 450nm would allow light with a wavelength less than 450nm to pass through the filter, but reflect light higher than 450nm.

Longpass Filter

By: Tim Bushnell, PhD

A filter that allows light over a set wavelength to pass through and reflects light below the set wavelength. For example, a longpass filter with a wavelength of 670nm would allow light with a wavelength greater than 670nm to pass through the filter, but reflect light lower than 670nm.

Electrons

By: Tim Bushnell, PhD

Counted by the Photomultiplier Tube (PMT) in the flow cytometer. Photons enter the PMT and the signal is amplified in the PMT when a photon strikes the anode and “knocks” of electrons. These electrons then hit a series of subsequent anodes, amplifying the total number of electrons of signal. The PMT then counts the total number of electrons and this is converted to the signal.